QTP4B071-16 governs qualification training for portable analysis and water sampling equipment in bioenvironmental engineering.
View QTP4B071-16 on armypubs.army.mil
It covers portable GC/MS, portable laboratory analysis kits, FT-IR, and water sampling equipment. The HAPSITE® criterion requires successful completion of all checklist items with no trainer assistance.
If the TIC is greater than 500,000 at the end of run, repeat clean-out. (paragraph 8)
All four criteria constitute a satisfactory blank run.
If there are additional peaks in the blank spectrum, and they cannot be removed with additional blank runs, review your blank chromatogram, note the additional analytes, and contact the ESOH Service Center. (paragraph 8)
The Blanking Reagent removes chlorine and chloramines from the water. (paragraph 9)
Measure the reacted sample 3-4 minutes after mixing the sample and reagents. (paragraph 7)