QTP4B051-16 governs noise calculations, noise surveys, audiometric booth surveys, and hearing protection adequacy.
View QTP4B051-16 on armypubs.army.mil
It contains task steps, review questions, performance checklists, and answers for noise calculations, source surveys, worker dosimetry, octave band surveys, audiometric booth surveys, and hearing protection device attenuation. It requires hands-on training and evaluation for the listed tasks, with specified levels of trainer assistance.
Select a type I sound level meter (SLM) with octave band analyzer and perform pre-calibration according to the manufacturer’s instructions. (paragraph 1)
Ensure booth testing conditions in relation to external noise/activity, door closed, etc., are representative of typical audiometric testing conditions. (paragraph 2)
Perform post-calibration of SLM according to the manufacturer’s instructions. (paragraph 7)
Random incidence microphones are designed for “grazing” angles; it must be hold at 70 degrees or nearly perpendicular to the source. (paragraph 2)
For 500 Hertz (Hz), < 27 dB. For 1,000 Hz, < 29 dB. For 2,000 Hz, < 34 dB. For 4,000 Hz, < 39 dB. For 8,000 Hz, < 41 dB. (paragraph 1)